1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 <2012> 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 <2012> 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 |
<== Date ==> | <== Thread ==> |
---|
Subject: | Re: Tektronix 3014 Tracing Problem |
From: | Pavel Maslov <[email protected]> |
To: | Eric Norum <[email protected]> |
Cc: | Yoojong Kim <[email protected]>, EPICS Tech Talk <[email protected]> |
Date: | Fri, 7 Dec 2012 05:38:53 +0100 |
It sounds like you have two problems. Generating an SDDS file is a separate issue from displaying waveforms in the MEDM window.
The fact that you see nothing in the plot region of the screen makes me think that there may be something wrong with your MEDM installation or the IOC that's acquiring the data.
To check the latter, you could try doing a 'camonitor' of one of the scope waveform records. Do the numbers look reasonable? Does the camonitor program produce a new waveform record each time that the scope triggers?
Have you ever been able to display a waveform record using the MEMD Cartesian Plot widget? Try creating a simple MEDM screen with a Cartesian Plot widget that gets its value from a scope waveform record and see if you can see the values.
Have you changed the waveform record names so that the MEDM widget can't find the data?
> <Tds3000 without a trace.png><Trace Issue.png><Trace10k.png>
On Dec 6, 2012, at 4:00 PM, Trevor Downer <[email protected]> wrote:
> Greetings and salutations!
>
> I am a student at Idaho State University working on a project to
> control a Tektronix 3014 oscilloscope, using EPICS, for the Idaho
> Accelerator Center.
> We have successfully installed EPICS and the necessary extensions to
> communicate with the TDS3014 oscilloscope. Using medm we cannot see
> the waveform tracings, we can however see the tracings from a web
> browser, however its refresh rate is quite slow.
> There seems to be a problem when I click on the SDDS Plot button, medm
> outputs a message in the terminal reading:
>
> "medm/trace10k: line 8: [: T: integer _expression_ expected
> Usage: trace [1-4][HV or LL]<CR>\n"
>
> Attached is a screen shot of the traceless oscilloscope in medm and
> the trace10k file where there seems to be a problem.
> I would greatly appreciate any help to fixing the tracing problem,
> thank you very much!
>
> -Trevor Downer
--
Eric Norum
[email protected]