1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 <2014> 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 <2014> 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 |
<== Date ==> | <== Thread ==> |
---|
Subject: | Re: read status nin "" 0 error with |
From: | Eric Norum <[email protected]> |
To: | Ana Malagon <[email protected]> |
Cc: | "[email protected]" <[email protected]> |
Date: | Wed, 10 Dec 2014 11:37:44 -0800 |
That seems reasonable, but the command ("FREQ:CW?”) doesn’t appear to be getting sent out to the instrument. Try turning on some more diagnostics asynSetTraceMask("L0", -1, 0xFF) to see everything. Also, you mentioned that you were unable to set the terminator character — this shouldn’t be a problem since I presume that the instrument uses the GPIB EOI line to mark the end of a message — but I do wonder why this doesn’t work. What error message do you see?
--
Eric Norum [email protected] |