1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 <2016> 2017 2018 2019 2020 2021 2022 2023 2024 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 <2016> 2017 2018 2019 2020 2021 2022 2023 2024 |
<== Date ==> | <== Thread ==> |
---|
Subject: | Re: Control system for Keithley |
From: | Márcio Paduan Donadio <[email protected]> |
To: | "[email protected]" <[email protected]> |
Date: | Wed, 28 Sep 2016 14:26:30 +0000 |
Dear Paramveer, So, the final goal will be to control your Keithley equipment using MEDM? Did I understand correctly? When you start Keithley IOC, can it connect correctly to the equipment? Before you try a GUI, I suggest you to read some PVs using caget / camonitor to see if it is tied to what you see on the equipment front screen. Also, it is a good idea to change some parameters with caput to see if the equipment reacts accordingly. Doing so you are sure that your IOC is working and communicating with Keithley. Next step is to test the same things in your GUI, that will work the same as camonitor and caput. If not, probably the PVs names are configured wrong in the GUI. Cheers, -- Márcio Paduan Donadio Engenheiro de Computação - Software de Operação das Linhas de Luz Laboratório Nacional de Luz Síncrotron – (LNLS) Centro Nacional de Pesquisa em Energia e Materiais (CNPEM) + 55 (19) 3512-3520 - [email protected] www.lnls.cnpem.br Em 28-09-2016 10:05, Paramveer Jain escreveu:
|