1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 <2024> 2025 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 <2024> 2025 |
<== Date ==> | <== Thread ==> |
---|
Subject: | Re: [EXTERNAL] Epics support for DIGITEL MPCQ Ion Pump Controller and Televac MX 200 Vacuum Controller |
From: | "Smith, Martin via Tech-talk" <tech-talk at aps.anl.gov> |
To: | Mrinal Bera <mrinalkb at uchicago.edu> |
Cc: | "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov> |
Date: | Fri, 23 Feb 2024 12:42:14 +0000 |
Here is what we use at APS for the Televac MX200 device. In your boot file you would include
one of these, here I show 2 different MX200 devices.
# For MM200 the Televac has two cold cathodes and a minimum of 2 convectrons
# if STN is either 5/6 then the corresponding CV1 are 1/2 and CV2 are 3/4
# if STN is either 3/4 then the corresponding CV is 1/2 and no CV2
# MM200 expects "\r" for EOS for both inputs and outputs.
#!dbLoadRecords("db/vs.db","P=FE:28:ID:,GAUGE=VS1,PORT=TV1,ADDR=0,DEV=MM200,STN=5")
#!dbLoadRecords("db/vs.db","P=FE:28:ID:,GAUGE=VS2,PORT=TV1,ADDR=0,DEV=MM200,STN=6")
#!dbLoadRecords("db/vs.db", "P=TV:,GAUGE=VS1,PORT=Televac1,ADDR=0,DEV=MX200,STN=3 1 0 1")
#!dbLoadRecords("db/vs.db", "P=TV:,GAUGE=VS2,PORT=Televac1,ADDR=0,DEV=MX200,STN=4 0 2 2")
Let me know if I left anything out.
Marty
From: Tech-talk <tech-talk-bounces at aps.anl.gov> on behalf of Mrinal Bera via Tech-talk <tech-talk at aps.anl.gov>
Sent: Thursday, February 22, 2024 5:04 PM To: Webb, Brad <webbsb at ornl.gov>; Abdalla.Ahmad at sesame.org.jo <Abdalla.Ahmad at sesame.org.jo> Cc: tech-talk at aps.anl.gov <tech-talk at aps.anl.gov> Subject: Re: [EXTERNAL] Epics support for DIGITEL MPCQ Ion Pump Controller and Televac MX 200 Vacuum Controller
This Message Is From an External Sender
This message came from outside your organization.
Hi Brad and Abdalla,
Thank you for your template and protocol files. I had to modify the files a little bit to make it work for MPCQ Pump. Best regards, Mrinal > On Feb 22, 2024, at 6:00 AM, Webb, Brad <webbsb at ornl.gov> wrote: > > Hi > > At SNS we use the DIGITEL QPC Controller from Gamma vacuum. > > We use the serial interface via a digi/moxa. > > Attached is our streamdevice interface (i.e., xxx.proto and > xxx.template) > > -Brad > > On Thu, 2024-02-22 at 00:50 +0000, Mrinal Bera via Tech-talk wrote: >> Did anybody write EPICS support for the following devices? 1) Gamma >> DIGITEL MPCQ Ion Pump >> Controller(https: //www. gammavacuum. com/products/digitel- >> controllers/3335/digitel-mpcq?number=MPCQ1PS1U1SSNI&c=17669) 2) >> Televac MX200 Vacuum Controller >> ZjQcmQRYFpfptBannerStart >> >> >> >> >> >> This Message Is From an External Sender >> >> >> >> This message came from outside your organization. >> >> >> >> >> >> >> >> ZjQcmQRYFpfptBannerEnd >> >> Did anybody write EPICS support for the following devices? >> 1) Gamma DIGITEL MPCQ Ion Pump >> Controller(https://www.gammavacuum.com/products/digitel-controllers/3335/digitel-mpcq?number=MPCQ1PS1U1SSNI&c=17669 >> ) >> 2) Televac MX200 Vacuum Controller >> (https://www.frederickscompany.com/products/mx200/) >> >> Thanks, >> Mrinal >> >> Mrinal Kanti Bera, PHD >> Beamline Scientist and Research Associate Professor >> NSF’s ChemMatCARS, Pritzker School of Molecular Engineering, >> The University of Chicago >> 9700 S. Cass Ave Bldg 434D >> Argonne, IL-60439 >> United States >> Phone: +1-630-252-0472 >> >> >> > |
Attachment:
mx200.tar
Description: mx200.tar