1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 <2012> 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 <2012> 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 |
<== Date ==> | <== Thread ==> |
---|
Subject: | Re: Tektronix 3014 Tracing Problem |
From: | "Hartman, Steven M." <[email protected]> |
To: | Yujong Kim <[email protected]> |
Cc: | EPICS Tech Talk <[email protected]> |
Date: | Fri, 7 Dec 2012 14:10:55 -0500 |
On Dec 6, 2012, at 11:51 PM, Yujong Kim wrote: > We can control all main functions of the scope from EPICS MEDM OPI panels without any problem. Hi Yujong-- To isolate whether this is an medm configuration issue or a CA configuration issue (e.g., EPICS_CA_MAX_ARRAY_BYTES), try to caget the waveform from the computer running medm. -- Steven Hartman [email protected]