1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 <2020> 2021 2022 2023 2024 2025 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 <2020> 2021 2022 2023 2024 2025 |
<== Date ==> | <== Thread ==> |
---|
Subject: | XSPRESS3 / InternalTrigger / AcquireTime |
From: | John Dobbins via Tech-talk <tech-talk at aps.anl.gov> |
To: | EPICS Tech Talk <tech-talk at aps.anl.gov> |
Date: | Wed, 26 Aug 2020 13:59:47 +0000 |
All,
FYI.
I noticed that on the xspress3 with internal trigger selected the AcquireTime rolls over for times greater than ~ 53 seconds.
The ADC sample time is 12.5 nS. It appears the AcquireTime is converted to a 32-bit integer of ADC samples. 12.5nS * 2^32 = 53.7 seconds.
Setting an AcquireTime of 60 seconds generates an exposure of ~ 7 seconds.
Apologies if this is reported elsewhere and I missed it.
John Dobbins
Research Support Specialist
Cornell High Energy Synchrotron Source
Cornell University
Maybe it's all part of a great big ineffable plan. - Neil Gaiman, _Good Omens_ |