1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024 
<== Date ==> <== Thread ==>

Subject: ADPilatus
From: John Dobbins via Tech-talk <tech-talk at aps.anl.gov>
To: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
Cc: "Arthur R. Woll" <arthurwoll at cornell.edu>
Date: Fri, 2 Oct 2020 14:29:55 +0000
All,

We have a number of DECTRIS PILATUS detectors for which I recently added an up-to date areaDetector installation.

Previously these detectors were using areaDetector R1-9-1, EPICS Base R3.14.11

I added areaDetector R3-9, EPICS Base R3.15.8

These are the three detector and their tvx versions - note that the 100K  tvx version is different.

PIL 100K   tvx-7.3.13.121212b

PIL 200K  tvx  tvxe-7.4.01-130212

PIL 300K   tvx  tvxe-7.4.01-130212

The 200K and 300K are working fine with the new software, however the 100K is not.

The image from the PIL 100K looks like this:

https://www.classe.cornell.edu/~jad/PILATUS_100K.PNG

The first 26 rows appear fine, i.e. pixels typically show tens of counts, then in row 27 pixels with very large numbers and pixels with zero start appearing.

Neither the IOC nor camserver is reporting errors.

The PILATUS 100K continues to work as expected with the old areaDetector installation.

Any ideas on how to proceed?

Regards,

John Dobbins

Research Support Specialist
Cornell High Energy Synchrotron Source
Cornell University











Replies:
RE: ADPilatus Mark Rivers via Tech-talk

Navigate by Date:
Prev: Re: Multiple Archiver Appliances on same host, different NICs Phillip A. Sorensen via Tech-talk
Next: RE: ADPilatus Mark Rivers via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024 
Navigate by Thread:
Prev: Re: Implementing closed loop on EPICS level Peterson, Kevin M. via Tech-talk
Next: RE: ADPilatus Mark Rivers via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024