1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 <2022> 2023 2024 | Index | 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 <2022> 2023 2024 |
<== Date ==> | <== Thread ==> |
---|
Subject: | ADEiger acquisiton problem |
From: | John Dobbins via Tech-talk <tech-talk at aps.anl.gov> |
To: | "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov> |
Date: | Wed, 13 Jul 2022 20:33:30 +0000 |
I am encountering a problem with acquisition on an EIGER2 which I haven't experienced before (on other EIGERs)
This appears to be true with both the latest ADEiger and an older version (Driver 2-5, ADCore 3.10.0)
I have an EIGER2 16M, firmware 2020.2.1
If I set TriggerMode = Internal Series, NumImages =1 , AcquireTime = 1.0 and press Start the MEDM screen reports the detector going through all the expected states (Arming, Triggering,
Processing Files) . An HDF5 master files appears but no data file. Also the image counter does not increment.
External Trigger works as expected.
When tested separately from EPICS (a locally written program sending HTTP requests) the acquisition occurs as expected.
Any advice appreciated.
John Dobbins
Research Support Specialist
Cornell High Energy Synchrotron Source
Cornell University
|