Experimental Physics and Industrial Control System
|
All,
FYI.
I noticed that on the xspress3 with internal trigger selected the AcquireTime rolls over for times greater than ~ 53 seconds.
The ADC sample time is 12.5 nS. It appears the AcquireTime is converted to a 32-bit integer of ADC samples. 12.5nS * 2^32 = 53.7 seconds.
Setting an AcquireTime of 60 seconds generates an exposure of ~ 7 seconds.
Apologies if this is reported elsewhere and I missed it.
John Dobbins
Research Support Specialist
Cornell High Energy Synchrotron Source
Cornell University
Maybe it's all part of a great big ineffable plan. - Neil Gaiman, _Good Omens_
|
- Replies:
- Re: XSPRESS3 / InternalTrigger / AcquireTime Matt Newville via Tech-talk
- Navigate by Date:
- Prev:
Re: ADViewers: EPICS_NTNDA_VIEWER' does not connect - ADViewers Mark Rivers via Tech-talk
- Next:
Re: XSPRESS3 / InternalTrigger / AcquireTime Matt Newville via Tech-talk
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
<2020>
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
asyn R4-40-1 available Mark Rivers via Tech-talk
- Next:
Re: XSPRESS3 / InternalTrigger / AcquireTime Matt Newville via Tech-talk
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
<2020>
2021
2022
2023
2024
|
ANJ, 26 Aug 2020 |
·
Home
·
News
·
About
·
Base
·
Modules
·
Extensions
·
Distributions
·
Download
·
·
Search
·
EPICS V4
·
IRMIS
·
Talk
·
Bugs
·
Documents
·
Links
·
Licensing
·
|