1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024 
<== Date ==> <== Thread ==>

Subject: Re: ADPilatus
From: John Dobbins via Tech-talk <tech-talk at aps.anl.gov>
To: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
Date: Tue, 6 Oct 2020 15:28:15 +0000
For the record:

The root of the problem appears to be that the new IOC contains a record (added in R2-7) for the x-ray energy, $(P)$(R)Energy which did not exist in the previous IOC.

The default value for this record is 20 KeV.

This value is used by camserver to generate a flat field correction file. On this Pilatus at least, the value 20 KeV generates a bad flat field correction file.

Thanks to Mark Rivers for his help in debugging this.

John Dobbins


 

 

From: Tech-talk <tech-talk-bounces at aps.anl.gov> On Behalf Of John Dobbins via Tech-talk
Sent: Friday, October 2, 2020 9:30 AM
To: tech-talk at aps.anl.gov
Cc: Arthur R. Woll <arthurwoll at cornell.edu>
Subject: ADPilatus

 

All,

 

We have a number of DECTRIS PILATUS detectors for which I recently added an up-to date areaDetector installation.

 

Previously these detectors were using areaDetector R1-9-1, EPICS Base R3.14.11

 

I added areaDetector R3-9, EPICS Base R3.15.8

 

These are the three detector and their tvx versions - note that the 100K  tvx version is different.

 

PIL 100K   tvx-7.3.13.121212b

 

PIL 200K  tvx  tvxe-7.4.01-130212

 

PIL 300K   tvx  tvxe-7.4.01-130212

 

The 200K and 300K are working fine with the new software, however the 100K is not.

 

The image from the PIL 100K looks like this:

 

 

The first 26 rows appear fine, i.e. pixels typically show tens of counts, then in row 27 pixels with very large numbers and pixels with zero start appearing.

 

Neither the IOC nor camserver is reporting errors.

 

The PILATUS 100K continues to work as expected with the old areaDetector installation.

 

Any ideas on how to proceed?

 

Regards,

 

John Dobbins

 

Research Support Specialist

Cornell High Energy Synchrotron Source

Cornell University

 

 

 

 

 

 

 

 

 

 


Replies:
RE: ADPilatus Mark Rivers via Tech-talk
References:
ADPilatus John Dobbins via Tech-talk
RE: ADPilatus Mark Rivers via Tech-talk
Re: ADPilatus John Dobbins via Tech-talk
RE: ADPilatus Mark Rivers via Tech-talk

Navigate by Date:
Prev: Re: Sequencer/SNL and PVA? Davide Marcato via Tech-talk
Next: RE: ADPilatus Mark Rivers via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024 
Navigate by Thread:
Prev: RE: ADPilatus Mark Rivers via Tech-talk
Next: RE: ADPilatus Mark Rivers via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  <20202021  2022  2023  2024 
ANJ, 06 Oct 2020 Valid HTML 4.01! · Home · News · About · Base · Modules · Extensions · Distributions · Download ·
· Search · EPICS V4 · IRMIS · Talk · Bugs · Documents · Links · Licensing ·