1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  <20222023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  <20222023  2024 
<== Date ==> <== Thread ==>

Subject: Only one network interface of three NF CA5351 Current Amplifiers is accessible
From: Chiang, Liang-Chih [江良志] via Tech-talk <tech-talk at aps.anl.gov>
To: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
Date: Tue, 22 Mar 2022 03:53:19 +0000

Hello,


I have three NF CA5351 Current Amplifiers and use Ethernet/LAN to access them remotely.

(product webpage: https://www.nfcorp.co.jp/english/pro/mi/loc/pre/ca5351/index.html  )


For some reason which I don't know, they would behave erroneously.

Only one of them is accessible remotely, and after several seconds, the accessible one will change.

That is,

only no. 1 is accessible,

 several seconds passed,

only no. 2 is accessible,

several seconds passed,

only no. 3 is accessible,

several seconds passed,

...repeat the above sequence...


I try rebooting the devices, doesn't work.


The solution I found is:

Unplug the network cable,

change the IP address of the device,

change the IP address back to the original setting,

plug the network cable.


Then everything is fine.


Does anyone have any idea about this bizarre behavior?



Best regards.



LiangChih Chiang

NSRRC Beamline group

email: chiang.lc at nsrrc.org.tw 


Replies:
RE: Only one network interface of three NF CA5351 Current Amplifiers is accessible Mark Rivers via Tech-talk

Navigate by Date:
Prev: Re: Unknown error with an equipment Mark Rivers via Tech-talk
Next: NDCodec Compression Factor Smith, William via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  <20222023  2024 
Navigate by Thread:
Prev: RE: Unknown error with an equipment Mark Rivers via Tech-talk
Next: RE: Only one network interface of three NF CA5351 Current Amplifiers is accessible Mark Rivers via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  <20222023  2024 
ANJ, 14 Sep 2022 Valid HTML 4.01! · Home · News · About · Base · Modules · Extensions · Distributions · Download ·
· Search · EPICS V4 · IRMIS · Talk · Bugs · Documents · Links · Licensing ·