Folks,
We use the CAENels TetrAMM electrometers for X-ray beam position and intensity measurements.
My units are running firmware 2.9.11. I see that more recent firmware supports a PKTSIZE command which allows control over the number of readings per Ethernet packet. That command would be useful to reduce buffering in the TetrAMM when
using it for fast feedback. I tried updating to firmware 3.3.03 today, but the update failed. Has anyone else tried such an update?
I also have a question about the newer firmware and newer TetrAMM manuals. V1.5 of the TetrAMM User Manual describes the GATE command, which allows collecting data only when a signal level is high. The EPICS quadEM driver supports that
command, and we use it with firmware 2.9.11. Versions 3.2 and 4.1 of that manual do not describe the GATE command. Does anyone know if that command has been eliminated in more recent firmware, or is it an oversight in the manual?
Thanks,
Mark
|
- Replies:
- Re: CAENels TetrAMM questions J. Lewis Muir via Tech-talk
- Navigate by Date:
- Prev:
RE: Make waveform from ai record Žiga Oven via Tech-talk
- Next:
Calcout from other IOC Gabriele Santambrogio via Tech-talk
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
<2025>
- Navigate by Thread:
- Prev:
RE: Make waveform from ai record Žiga Oven via Tech-talk
- Next:
Re: CAENels TetrAMM questions J. Lewis Muir via Tech-talk
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
<2025>
|
ANJ, 25 Feb 2025 |
·
Home
·
News
·
About
·
Base
·
Modules
·
Extensions
·
Distributions
·
·
Download
·
Search
·
IRMIS
·
Talk
·
Documents
·
Links
·
Licensing
·
|