1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 
<== Date ==> <== Thread ==>

Subject: RE: [External] Support for Edwards nXDS 10i
From: "Leblanc, Gregory via Tech-talk" <[email protected]>
To: Heinz Junkes <[email protected]>, "[email protected]" <[email protected]>
Date: Wed, 16 Jul 2025 20:19:10 +0000
Heinz,

Do you have the correct pins tied together in the 15 pin D-shell to enable serial?  I think for RS-232 you need to pull pin 5 low, which they suggest doing by tying to pin 2.  For RS-485, you also need to pull pin 6 low.  Caveat: I haven't tried this yet, but I did have this in my notes as part of the next project.
     Greg

P.S. My instruction manual is supposedly A73501860 Issue B

--
Gregory Leblanc
Accelerator Engineer
Edwards Accelerator Lab - Ohio University
123 University Terrace
Athens, OH 45701 USA
[email protected]
M: (401) 52-OUAL1 or (401) 526-8251
https://urldefense.us/v3/__https://orcid.org/0000-0001-7899-5296__;!!G_uCfscf7eWS!bpWEVp-xmC0vSim4sEOLLS6FPKniiQHRMRWllRnadNk5YCspf4gXlElM6QHcRa8xke-ZBiUxtYkHwNEpE6-KnQ$ 

-----Original Message-----
From: Tech-talk <[email protected]> On Behalf Of Heinz Junkes via Tech-talk
Sent: Wednesday, July 16, 2025 6:41 AM
To: [email protected]
Subject: [External] Support for Edwards nXDS 10i

Hi Techies,

We are trying to commission an Edwards scroll pump nXDS 10i with streamdevice via serial communication and streamdevice.
We have a protocol description with serial 9600 baud and Ascii communication like "?V808\cr”

But unfortunately with these settings nothing happens on the link.

There is a Windows-Programm whitch runs with 38400 baud and binary communication protocol ( named Modbus?).
This works.

Does anyone have such a pump in operation?
Are there different protocol variants?
Do we have the wrong documentation?

Danke Heinz


Replies:
Re: [External] Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
References:
Support for Edwards nXDS 10i Heinz Junkes via Tech-talk

Navigate by Date:
Prev: Re: Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
Next: Re: [External] Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 
Navigate by Thread:
Prev: Re: Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
Next: Re: [External] Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 