1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 
<== Date ==> <== Thread ==>

Subject: Re: [External] Support for Edwards nXDS 10i
From: Heinz Junkes via Tech-talk <[email protected]>
To: "Leblanc, Gregory" <[email protected]>
Cc: "[email protected]" <[email protected]>
Date: Wed, 16 Jul 2025 22:51:20 +0200
Hallo Greg,
Yes, we did exactly the same and followed the instructions from "A73501860 Issue B" on page 5.
We have now checked several times that this is really the case.
Unfortunately, the system does not behave as described :-(

Danke Heinz

> On 16. Jul 2025, at 22:19, Leblanc, Gregory <[email protected]> wrote:
>
> Heinz,
>
> Do you have the correct pins tied together in the 15 pin D-shell to enable serial?  I think for RS-232 you need to pull pin 5 low, which they suggest doing by tying to pin 2.  For RS-485, you also need to pull pin 6 low.  Caveat: I haven't tried this yet, but I did have this in my notes as part of the next project.
>     Greg
>
> P.S. My instruction manual is supposedly A73501860 Issue B
>
> --
> Gregory Leblanc
> Accelerator Engineer
> Edwards Accelerator Lab - Ohio University
> 123 University Terrace
> Athens, OH 45701 USA
> [email protected]
> M: (401) 52-OUAL1 or (401) 526-8251
> https://orcid.org/0000-0001-7899-5296
>
> -----Original Message-----
> From: Tech-talk <[email protected]> On Behalf Of Heinz Junkes via Tech-talk
> Sent: Wednesday, July 16, 2025 6:41 AM
> To: [email protected]
> Subject: [External] Support for Edwards nXDS 10i
>
> Hi Techies,
>
> We are trying to commission an Edwards scroll pump nXDS 10i with streamdevice via serial communication and streamdevice.
> We have a protocol description with serial 9600 baud and Ascii communication like "?V808\cr”
>
> But unfortunately with these settings nothing happens on the link.
>
> There is a Windows-Programm whitch runs with 38400 baud and binary communication protocol ( named Modbus?).
> This works.
>
> Does anyone have such a pump in operation?
> Are there different protocol variants?
> Do we have the wrong documentation?
>
> Danke Heinz
>

Attachment: smime.p7s
Description: S/MIME cryptographic signature


References:
Support for Edwards nXDS 10i Heinz Junkes via Tech-talk
RE: [External] Support for Edwards nXDS 10i Leblanc, Gregory via Tech-talk

Navigate by Date:
Prev: RE: [External] Support for Edwards nXDS 10i Leblanc, Gregory via Tech-talk
Next: documentation for new epics 7 dbLock usage and Redundant IOC software Baily, Scott A via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 
Navigate by Thread:
Prev: RE: [External] Support for Edwards nXDS 10i Leblanc, Gregory via Tech-talk
Next: documentation for new epics 7 dbLock usage and Redundant IOC software Baily, Scott A via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  2021  2022  2023  2024  <20252026 
ANJ, 19 Mar 2026 · Home · News · About · Talk · Base · Modules · Extensions ·
· Distributions · Download · Documents · Links · Licensing ·