Hi Michael,
We’ve used 4-wire PT100 RTDs with the Lakeshore 224, which has support for 12 channels. It’s easy to interface to and setup.
https://www.lakeshore.com/products/categories/overview/temperature-products/cryogenic-temperature-monitors/model-224-temperature-monitor
We use it for both cryogenic applications and also room temperature measurements.
It has good EPICS support, although I’m not seeing any published support on the EPICS website. I can send you a copy of our local support if you need it.
Cheers,
Matt
From: Tech-talk <[email protected]>
On Behalf Of Sintschuk, Michael via Tech-talk
Sent: Friday, August 22, 2025 5:11 AM
To: tech-talk <[email protected]>
Subject: [EXTERNAL] 4-wire sensing device
Dear all, I would like to know what devices you are using for measuring 4-wire RTDs (temperatur), that -of course- have EPICS-support. For a long
time, we used the multimeter Keithley K2700, until it broke. The Keithley DAQ6510/7700 seems to
Dear all,
I would like to know what devices you are using for measuring 4-wire RTDs (temperatur), that -of course- have EPICS-support.
For a long time, we used the multimeter Keithley K2700, until it broke. The Keithley DAQ6510/7700 seems to be the
successor of the K2700 that would fit our needs. But I wonder if there are other systems in use…
We don’t actually need high measurement accuracy, but at least 10 channels would be good.
Best regards
Michael