Eric,
The two values I seem to have access to via telnet to the gateway are:
Lan Timeout : 10 (seconds)
IO-Timoute : 120 (seconds)
Eric Norum wrote:
>
> On Wednesday, August 21, 2002, at 11:53 AM, [email protected] wrote:
>
> > I am using the HP (Agilent) E2050A Lan /GPIB gateway with Epics 3.14b2.
> > I have some device support for a scope which I've modified from pre
> > 3.14 - and it's running on a solaris IOC. If I leave the device (the
> > gateway) unaddressed for a 'long' period of time (~30 minutes) and then
> > send a command to a GPIB instrument, I receive the following message:
> >
> > hpWriteCmd(142f28, 1320b4, 1): RPC: Unable to receive; An event
> > requires attention
> >
> > At this point, the only way to re-establish communications is to
> > "reboot" the ioc.
> >
> > Any ideas here?
> >
> >
>
> What values do you have the 2050A communication timeouts set to?
>
> --
> Eric Norum <[email protected]>
> Department of Electrical Engineering
> University of Saskatchewan
> Saskatoon, Canada.
> Phone: (306) 966-5394 FAX: (306) 966-5407
--
+-----------------------------------------+
| S. Joshua Stein |
| |
| Electronics Engineer |
| Accelerator Systems Division |
| Advanced Photon Source |
| Argonne National Laboratory |
| www.aps.anl.gov/asd/people/stein/htdocs |
+-----------------------------------------+
- References:
- Re: HP Lan-GPIB gateway Eric Norum
- Navigate by Date:
- Prev:
Re: HP Lan-GPIB gateway Eric Norum
- Next:
Re: Siemens/Profibus Drivers Matthias Clausen
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
<2002>
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
- Navigate by Thread:
- Prev:
Re: HP Lan-GPIB gateway Eric Norum
- Next:
timestamp question Kevin Tsubota
- Index:
1994
1995
1996
1997
1998
1999
2000
2001
<2002>
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
|