EPICS Controls Argonne National Laboratory

Experimental Physics and
Industrial Control System

1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  <20212022  2023  2024  Index 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  <20212022  2023  2024 
<== Date ==> <== Thread ==>

Subject: Re: Re: Re: detector state error and writing status error for marCCD
From: 秦文明 via Tech-talk <tech-talk at aps.anl.gov>
To: "Mark Rivers" <rivers at cars.uchicago.edu>
Cc: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
Date: Thu, 22 Jul 2021 15:09:37 +0800 (GMT+08:00)
Dear Mark,

Sorry for late reply. The fiber-optical was broken. We haven't got time to do the test. We are running the IOC on another machine, not on the marccd software itself. This marCCD is a spare detector which haven't been use for last year. We are doing the test for future use. We will test it again after the fiber problem been solved. Thanks a lot.

Best Regards,

Wenming


&gt; -----原始邮件-----
&gt; 发件人: "Mark Rivers" <rivers at cars.uchicago.edu>
&gt; 发送时间: 2021-07-16 20:14:13 (星期五)
&gt; 收件人: "秦文明" <qinwenming at sari.ac.cn>
&gt; 抄送: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
&gt; 主题: Re: Re: detector state error and writing status error for marCCD
&gt; 
&gt; What machine are you running the IOC on?  Is it the machine that is running the marccd software itself, or is it some other machine?
&gt; 
&gt; 
&gt; Have you tried saving a file using the marccd software itself, i.e. without using EPICS, to see if that works?
&gt; 
&gt; 
&gt; The marccd server logs all of the activity when it communicates with EPICS.  Have you looked at those messages for any errors?
&gt; 
&gt; 
&gt; Mark
&gt; 
&gt; 
&gt; 
&gt; ________________________________
&gt; From: 秦文明 <qinwenming at sari.ac.cn>
&gt; Sent: Friday, July 16, 2021 1:41 AM
&gt; To: Mark Rivers
&gt; Cc: tech-talk at aps.anl.gov
&gt; Subject: Re: Re: detector state error and writing status error for marCCD
&gt; 
&gt; Dear Mark,
&gt; 
&gt; Thank you very much for your reply. I tried to collect with /home/marccd, and /home/marccd/Documents, It still didn't work. The /home/marccd/Documents should have write permission, because we had written data in this fold before with BluIce.
&gt; 
&gt; Best Regards,
&gt; 
&gt; Wenming
&gt; 
&gt; 
&gt; &gt; -----原始邮件-----
&gt; &gt; 发件人: "Mark Rivers" <rivers at cars.uchicago.edu>
&gt; &gt; 发送时间: 2021-07-13 20:01:55 (星期二)
&gt; &gt; 收件人: "秦文明" <qinwenming at sari.ac.cn>
&gt; &gt; 抄送: "tech-talk at aps.anl.gov" <tech-talk at aps.anl.gov>
&gt; &gt; 主题: Re: detector state error and writing status error for marCCD
&gt; &gt;
&gt; &gt;
&gt; &gt; Hi Wenming,
&gt; &gt;
&gt; &gt;
&gt; &gt; You are telling marccd to save files to /home/.  I suspect that the marccd account does not have write permission to /home/.  Try writing to /home/marccd, or some other directory where marccd does have write permission.
&gt; &gt;
&gt; &gt;
&gt; &gt; Mark
&gt; &gt;
&gt; &gt;
&gt; &gt;
&gt; &gt; ________________________________
&gt; &gt; From: Tech-talk <tech-talk-bounces at aps.anl.gov> on behalf of 秦文明 via Tech-talk <tech-talk at aps.anl.gov>
&gt; &gt; Sent: Monday, July 12, 2021 8:10 PM
&gt; &gt; To: tech-talk at aps.anl.gov
&gt; &gt; Subject: detector state error and writing status error for marCCD
&gt; &gt;
&gt; &gt;
&gt; &gt; Dear Sir/Madam,
&gt; &gt;
&gt; &gt;
&gt; &gt; I installed and configured the ADmarCCD module on CentOS7.  The installation looks succesed. But it always shows detector state error and writing status error in MEDM gui as shown in the following picture. Would you please give me some suggestion on how to figure out it. Thanks a lot.
&gt; &gt;
&gt; &gt;
&gt; &gt; [cid:26b09179$1$17a9d695785$Coremail$qinwenming$sari.ac.cn]
&gt; &gt;
&gt; &gt;
&gt; &gt;
&gt; &gt;
&gt; &gt;
&gt; &gt; Best Regards,
&gt; &gt;
&gt; &gt;
&gt; &gt; Wenming
&gt; </tech-talk at aps.anl.gov></tech-talk-bounces at aps.anl.gov></tech-talk at aps.anl.gov></qinwenming at sari.ac.cn></rivers at cars.uchicago.edu>
</qinwenming at sari.ac.cn></tech-talk at aps.anl.gov></qinwenming at sari.ac.cn></rivers at cars.uchicago.edu>

References:
detector state error and writing status error for marCCD 秦文明 via Tech-talk
Re: detector state error and writing status error for marCCD Mark Rivers via Tech-talk
Re: Re: detector state error and writing status error for marCCD 秦文明 via Tech-talk
Re: Re: detector state error and writing status error for marCCD Mark Rivers via Tech-talk

Navigate by Date:
Prev: Re: EPICS compilation for gcc and clang Philipp Kampmann via Tech-talk
Next: pyepcis does not seem to honor EPICS_CA_ADDR_LIST Koennecke Mark (PSI) via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  <20212022  2023  2024 
Navigate by Thread:
Prev: Re: Re: detector state error and writing status error for marCCD Mark Rivers via Tech-talk
Next: Building autosave with EPICS 7.0.6 and RTEMS5 Matt Rippa via Tech-talk
Index: 1994  1995  1996  1997  1998  1999  2000  2001  2002  2003  2004  2005  2006  2007  2008  2009  2010  2011  2012  2013  2014  2015  2016  2017  2018  2019  2020  <20212022  2023  2024 
ANJ, 22 Jul 2021 Valid HTML 4.01! · Home · News · About · Base · Modules · Extensions · Distributions · Download ·
· Search · EPICS V4 · IRMIS · Talk · Bugs · Documents · Links · Licensing ·